High-Speed Cell Sorting XYZ System | automated isolation of cells of complex samples
3D Print / Dispensing Systems

782300:176.26

High speed automated cell sorting

A leading manufacturer of measuring instruments and test arrays in the field of clinical diagnostics and life science required an innovative positioning solution for Fluorescence activated cell sorting of live cells. For this requirement, we developed a multi-axis solution to simplify the most complex sorting tasks for high-speed flow cytometric sorters.

This XYZ System sorting cells to 500 nm:

  • High-speed motorization ensures maximum sorting speed
  • Ultra-sensitive positioning with 500 nm resolution for identification and separation of particles below 1µm
  • Rigid mechanics for Stable fluidic design and precise droplet formation millions of times
  • Maximum dynamics and short process times with speeds up to 1,000 mm/s in X and Y
  • Cleanroom compatible with extremely long lifetime
  • For heavy duty automated 24/7 production

 

Prepared for sterile technology and innovative motion systems

This high-speed XYZ cell positioning system impresses with speed up to 530 millimeters per second and reduced complexity. The motion resolution of 0.5µm at high speed enables you to quickly detect cells and particles at sub-µm level and then collect them in separate collection vessels. You avoid contamination by highly accurate mechanics for optimized droplet break-off point.

Further, a new system with moving mass compensation was developed for sorting speed with movements up to 1000mm/s. No matter whether you are researching cancer cells, microparticles or microorganisms - this XYZ positioning system is individually customizable to your application and sample complexity.

In conventional cell sorters, there is a three-way trade-off between sort speed, sort purity and sort yield. This XYZ Stage allows selecting for purity in one sort-direction and yield in another, enabling simultaneous recovery of virtually all the target cells that would otherwise be lost to achieve that high purity. This enables you to sort faster without losing extremely valuable cells, and without sacrificing purity of the principal sort.

A fully interpolated controller guarantees fast 3D movements. The positioning system prevents the penetration of liquids and emission of particles with closed cover of all axes throughout the entire movement. The sterile compatible design is guaranteed by smooth, wipeable and metallic surfaces. In addition, the system is capable of logging, storing and outputting all operating and control parameters in real time for each batch and thus fully meets the high documentation requirements of medical technology.

 

Individual adaptation to your application and the complexity of the sample

  • Individual 3D design, customized to your concept
  • Adaptable with up to five powerful secondary Z-axes for setting up sorting or printing assemblies for cell samples, structures and organs and for printing individual lab-on-chip trays and for isolating cells in cancer therapy
  • Available with fully interpolated controller for fast 3D motion in space
  • Life cycle tests, prototype construction, functional samples, documentation and installation service
  • In-house precision special part production for customized requirements
  • On request completion to the complete system
  • Series production of your machine up to 500 pieces per year

    782300:176.26.2

    X

    Y

    Zp

    (Zs1 … 5)

    Travel

    [mm; deg]

    300-500

    200-500

    100

    100

    Repeatability unidirectional

    [µm; deg]

    ± 0.5

    ± 0.5

    ± 2.5

    ± 2.5

    Repeatability bidirectional

    [µm; deg]

    ± 0.4

    ± 0.4

    ± 2

    ± 2

    Positioning speed

    [mm/s; deg/s]

    530

    530

    130

    100

    Max. Load

    [N]

    200

    100

    300

    80

    Motor

     

    Iron Core Dynamic Linear Motor

    Iron Core Dynamic Linear Motor

    AC-Servo

    AC-Servo

    Drive

     

    Ball Screw

    Ball Screw

    Feedback

    Linear Scale

    Linear Scale

    Linear Scale

    Linear Scale

     

    Subject to change. Values ​​apply to single axes with our controllers.